JYTOP MS100 SpO2 Pulse Rate Blood Oxygen Simulator Pulse Oximeter reaction time
JYTOP MS100 SpO2 Pulse Rate Blood Oxygen Simulator Pulse Oximeter reaction time
Brief Introduction
MS100 SpO2 Simulator is a kind of Separated SpO2 simulator, small and light. It can perform a series of tests for the oximeter by simulation means, and gives cognizance of veracity about the oximeter.
Because the oximeter manufacturer may use different R-curves, some R-curves used abroad currently have been embedded.
Features
◆ Separated connection between simulator probe and host, operation and test are more convenient.
◆ 262K color and 320*240 TFT, adjustable brightness level.
◆ Film key-press, makes the operation more comfortable.
◆ Power with chargeable lithium battery, can display the charge information of battery.
◆ Key volume, can open or close.
Main Functions
◆ Oxygen saturation simulation;
◆ Pulse rate simulation;
◆ Patient state simulation preset;
◆ Can test the reaction time of device;
◆ Can simulate the SpO2 and pulse rate under different amplitude;
◆ Can test the performance of device under different interference source;
◆ Can select different R-curve.
Main Parameters
◆ SpO2
Measuring range: 35%~100%
Resolution:1%
Error: When SpO2 is 75%~100%, the error is the larger between ±2% and ±P(P denotes the precision of the tested oximeter); when SpO2 is 74%~50%, the error is the larger between ±2% and ±P; no define when SpO2 is less than 50%.
◆ Pulse rate
Measuring range:25bpm~250bpm
Resolution:5bpm
Error:1%±1bpm
◆ Amplitude
Measuring range:0.000%~20.000%
Resolution:1% for 1.000%~20.000%; 0.1% for 0.100%~0.900%; 0.025% for 0.000%~0.075%.
◆ Patient state
24 conditions presetting and 8 conditions as default after start. The number of condition can be adjusted by Motion level setup.
◆ Voltage
DC 3.6V~4.2V
◆ Power Supply
Voltage 3.7 rechargeable lithium battery × 1
◆ Battery working life
Charge and discharge no less than 500 times.
Accessaries
◆ A MS100 user manual
◆ A 5.0V adaptor