JYTOP MS400 Multiparameter Simulator multi-parameter Color Touch patient monitor

Sonderpreis $448.00 Normaler Preis $800.00

Introduction

MS400 Multiparameter Simulator is a portable device, features in small volume and light weight. It can simulate accurately 12-lead ECG, Respiration, Temperature, 2-channel IBP or 4-channel IBP. And it can be tested in where the monitor is used.

Features

  1. Simulate 35 kinds of arrhythmia (including Atrial, Ventricular, Premature Beat and Conduction Defects) and pacemaker waveforms (including Asynchronous, Demand Occ Sinus, Demand Freq Sinus, A/V Sequential, Non Capture, Non Function).
  2. Simulate four-channel IBP data.
  3. Simulate respiration signal with adjustable respiration rate, impedance and baseline impedance, and three kinds of apnea data.
  4. Simulate four kinds of temperature data.
  5. Simulate square wave, triangle wave, pulse wave and sine wave with adjustable frequency.
  6. Simulate standard ECG calibration, adult and pediatric ECG signal.
  7. ECG waveform complied with the standard of ANSI/AAMI EC13-2002, AHA, MIT ECG database, other ECG database and user-defined waveform data stored in TF card can be displayed by the simulator and exported to the equipments to be tested.
  8. Each waveform can be displayed by sliding the 3.5'' TFT touch screen, easy to operate.
  9. Power supply: rechargeable lithium battery or linear adapter, which reduces interference.
  10. Optional operation modes: keys and touch screen.

 

Parameters:

 

  • ECG
    Conventional: All leads have the same voltage with the II-lead as the reference
    Amplitude: 0.5,1.0,1.5 and 2.0mV
    ECG Rate: 15,20,25,30,40,60,80,100,120,140,160,180,200,220,240,260,280,300,320,340 and 350 BPM
    ECG artifacts: 50 and 60 Hz, muscle, baseline and breathing
    The elevation of the ST segment: +0.8 to -0.8 mV, the step is 0.1 mV
  • Arrhythmia
    Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia, Missed Beat, Atrial Tachycardia, Nodal Rhythm, Supraventricular Tachycardia
    Ventricular: Vent Fib (coarse), Vent Fib (fine), PVCs 6/min, PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy, Paired PVCs, Run 5 PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
    Premature Beat: Atrial PAC, Nodal PNC, PVC1 LV Focus, PVC1 E LV Focus, R on T LV Focus, PVC2 RV Focus, PVC2 E RV Focus, R on T RV Focus, Multifocal PVC
    Conduction Defects: First Degree Heart Block, Second Degree Heart Block, Third Degree Heart Block, Right Bundle Branch Block, Left Bundle Branch Block
  • Four IBP Channels
    The default: 00/00mmHg
    Dynamic IBP(unit: mmHg)
    Channel 1:

Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Ventricle: 25/00
Channel 2:

Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Atrium(CVP): 15/10
Right Ventricle: 25/00
Pulmonary Artery: 25/10
Pulmonary Wedge: 10/2
Left Atrium: 14/4
Channel 3:

Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Atrium(CVP): 15/10
Right Ventricle: 25/00
Pulmonary Artery: 25/10
Pulmonary Wedge: 10/2
Left Atrium: 14/4
Channel 4:
Auto Swan-Ganz (every 15 seconds): RA(CVP), RV, PA, PAW (Pulmonary Wedge)
Man Swan-Ganz: press ENTER key to change every time
Static-Pressure Levels
Channel 1: 0, 80, 160, 240, 320 and 400mmHg
Channel 2: 0, 50, 100, 150, 200 and 250mmHg
Channel 3: 0, 20, 40, 60, 80 and 100mmHg
Channel 4: P4 tracks the static pressure levels of P3

  • Respiration
    Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIN
    Impedance: 0.2, 0.5, 1.0 and 3.0 Ω
    Baseline impedance: 500, 1000, 1500, 2000Ω
    Apnea: 12 s, 22 s, 32 s
  • Temperature
    Four temperature: 0℃(32°F), 24℃( 75°F), 37℃(98.6°F), 40℃(104°F)

Accessories

Standard accessories 9V/1A(CE) power adapter, an User manual, a packaging box, a plastic packaging.